INTENDED USE

This Anti-HCV ELISA is a qualitative enzyme immunoassay for the in vitro detection of antibodies to Hepatitis C Virus(HCV) in human serum or plasma. It is intended for screening of blood donors and diagnosing patients related to infection with Hepatitis C Virus.

ASSAY PROCEDURE

  1. Prepare Reagents: Dilute 1 volume of Concentrated Washing Buffer (20×)with 19 volumes of distilled water,mix well.
  1. Add Samples: Open the foil pouch and remove the Microplate. Set up 1 well as Blank, 2 wells as negative control, 2 wells as positive control. After dispensing 100µL of Sample Diluent , dispense 10µL of sample or negative control or positive control to the respective wells. Gently vibrating the plate.
  1. Incubate: Cover the Microplate with plate cover and incubate the Microplate in a thermostat-controlled water-bath or microplate incubator at 37℃ for 30 minutes.
  1. Wash the Plate: Remove the plate cover. Aspirate the contents of all wells. Fill the wells with the diluted washing buffer ( 10~20 seconds to soak) then aspirate again. Repeat the procedure for 5 times. Make sure that the rest volume is minimal, by tapping plate onto absorbent paper.
  1. Add Conjugate: Add 100µL of conjugate to each well (except the blank well).
  2. Incubate: Cover the Microplate and incubate the plate at 37℃ for 20 minutes.
  3. Wash the Plate: Repeat the wash procedure as in step 4.
  4. Add Substrate: Add 50µL of Substrate Solution A and 50µL of Substrate Solution B to each well, mix well. Cover and incubate at 37℃ for 10 minutes.
  1. Stop reaction: Add 50µL Stop Solution to each well, mix well.
  2. Read the absorbance at 450 nm. If a dual wavelength measurement is used, the reference wavelength should be selected from 620nm to 690nm

RESULTS

  1. For the assay to be valid, the mean OD value of negative controls must be less than or equal to 0.1 and the mean OD value of positive controls must be greater than or equal to 0.8.
  1. Cut off value = 0.1+NC

NC= the mean absorbance value for two negative controls

     Important: If the NC is lower than 0.05, take it as 0.05.

  1. OD value of the sample ≥ cut off value, it is positive for anti-HCV. OD value of the sample < cut off value, it is negative for anti-HCV.

PRECAUTIONS

  1. Allow all kit components to reach room temperature before use.
  2. Follow the direction insert to control the reaction temperature and time strictly.
  3. Do not mix components of different lot numbers to use.
  4. The kit should be store at 2-8℃. Do not use kit components beyond their expiration date.

Specification:96T